Evaluation of Corneal Biomechanical Properties Measured By Ocular Response Analyzer in Cases with Rheumatoid Arthritis
PDF

Original Article

Evaluation of Corneal Biomechanical Properties Measured By Ocular Response Analyzer in Cases with Rheumatoid Arthritis

Turk J Ophthalmol 2011;41(3):167-170 DOI: 10.4274/tjo.41.09797
, , , , , ,
PDF
Permissions

Abstract

Purpose:

To compare the corneal biomechanical properties using an ocular response analyzer (ORA) in patients with rheumatoid arthritis and healthy control subjects.

Material and Method:

Thirty four eyes of 17 healthy individuals (control group) and 34 eyes of 17 patients with RA (study group) were recruited. ORA measurements were performed in all patients. Corneal-compensated intraocular pressure (IOPcc), Goldman related intraocular pressure (IOPg), corneal resistance factor (CRF), and corneal hysteresis (CH) were recorded. Statistical analyses were performed using a t test for independent samples.

Results:

In the study group, mean CH was 11.5±2.1 (7.7-16.2) mmHg, mean CRF was 12.1±2.3 (8.2-17) mmHg, IOPg 17.2±4.4 (8.3-25.1) and mean IOPcc was 16.5±4.3 (7.7-24.2) mmHg. In the control group, mean CH was 11.6±1.1 (9.7-13.8) mmHg, mean CRF was (11.8±1.2 (9.8-14.1) mmHg, IOPg 15.9±3.3 (9.4-25.7), and mean IOPcc was 15.4±2.6 (8.8-21.2) mmHg. There were no significant differences in these parameters between the study and control groups (p>0.05).

Discussion:

The biomechanical parameters measured with ORA were not statistical difference in patients with rheumatoid arthritis and age matched healthy individuals. (Turk J Ophthalmol 2011; 41: 167-70)

Keywords: Rheumatoid arthritis, Ocular response analyzer, Corneal hysteresis